Widely used in electric power, telecommunications, meteorology, oil fields, construction and industrial electrical equipment, such as grounding resistance measurement, circuit resistance measurement.
Note:
Regarding product purchase, technical maintenance and other services, welcome to send email:
ETCR2000 Clamp Type Earth Resistance Tester is a major breakthrough in the traditional grounding resistance measurement technology, which is widely used in the grounding resistance measurement of electric power, telecommunications, meteorology, oil fields, construction and industrial electrical equipment.
The ETCR2000 Clamp Type Earth Resistance Tester does not need to disconnect the grounding lead and auxiliary electrode when measuring the loaned Ground system, which is safe, fast and easy to use.
ETCR2000 Clamp Type Earth Resistance Tester can measure the Ground fault that cannot be measured by traditional methods, and it can be applied to the situation that cannot be measured by traditional methods, because ETCR2000 Clamp Type Earth Resistance Tester measures the comprehensive value of the resistance of grounding body and the resistance of grounding lead.
In addition, the C - type clamp ground resistance meter can also measure the leakage current and neutral current of the ground system.
Product nickname:Single jaw grounding resistance tester, nipper plier type grounding resistance tester, grounding resistance meter, pliers grounding resistance meter, clip-on grounding resistance meter, clip-on grounding resistance meter, grounding resistance, surface, grounding resistance tester, grounding resistance tester, grounding resistance measuring instrument, clamp type digital grounding resistance tester, grounding resistance tester, pincer pliers type grounding resistance measuring instrument
power supply | 6VDC(4 no.5 alkaline dry batteries) |
operating temperature | -10℃-55℃ |
Relative humidity | 10%-90% |
LCD | 4-digit LCD digital display, length and width 47×28.5mm |
clamp table weight (including battery) | 1130g |
clamp table size | length 275mm, width 85mm, thickness 56mm |
protection level | double insulation |
structure characteristics | clamping mode |
range shift | Automatic |
external magnetic field | <40A/m |
External electric field | <1V/m |
single measurement time | 1 SEC |
resistance measurement frequency | >1KHz |
highest resolution resistance measurement | 0.001Ω |
Resistance measurement range | 0.01-1000Ω |
* current measurement range | 0-30A(type C) |
measured current frequency | 50/60Hz |
stored measurement data | 99units |
setting range of resistance alarm critical value | 1-199Ω |
* setting range of current alarm critical value | 1-499mA(type C) |
1.The appearance of the product shall be subject to the actual product,Technical parameters with model if there are any changes,without notice.In case of misappropriation or copying, legal liability will be investigated.
2. For more information about Wuhan UHV, please enter product center . Service email:qiao@hvtest.cc
Enterprise products in strict accordance with international and national standards and rules of industry production, through the national, provincial and municipal electric power research institute and electric power authorities detection, measuring center, fully passed the ISO9001 international quality management system certification, the licence for manufacturing measuring instruments of the People's Republic of China, GB/T19001-2016 quality management system requirements, GB/T24001-2016 environmental management system requirements, GB/T45001-2020 occupational health and safety management system requirements.