| Equipment operating environment |
temperature:5℃~40℃ |
relative humidity:<80%(25℃时) |
| altitude:<2500m |
| mains frequency:50Hz±0.5Hz |
mains input:220V±5V |
| range of measurement |
cophase component(%):0.0001~200.0 |
distinguishability:0.0001 |
| orthogonal component(part):0.001~700.0 |
distinguishability:0.001 |
| impedance(Ω):0.0001~20.0 |
distinguishability:0.0001 |
| admittance(ms):0.0001~20.0 |
distinguishability:0.0001 |
| intrinsic error |
cophase component |
ΔX=±(X×2%+Y×2%)±Dx(Level 1 is optional) |
| orthogonal component |
ΔY=±(X×2%+Y×2%)±Dy(Level 1 is optional) |
| "X"、"Y"——Display value of instrument |
| "Dx,Dy"——Display value of instrument |
| Dx=2,Dy=5 |
| dial indicator |
PETS2(Level 1 is optional) |
| working range |
electricity |
(1%~149%)In (In =5A) |
| (5%~149%)In (In =1A) |
| voltage |
(5%~149%)Un (Un =100V,150V,100V/) |
| (5%~149%)Un (Un =100V/3) |
| workload |
electricity |
TO to TX<0.12Ω cosΦ=1 |
| voltage |
a to x<0.25VA (100V) |
| Polarity error indication |
When the operating current (voltage) reaches more than 5% of the rated current (voltage) and the error exceeds 180%, there should be polarity indication. |
| Note: if it is more than 10% of the rated operating current (voltage), there is still no proper polarity indication, indicating a fault. Please do not increase the current (voltage) to avoid burning out the instrument. |
| Variable ratio error indication |
When the operating current (voltage) reaches more than 5% of the rated current (voltage) and the error exceeds 30% but is less than 180%, the error indication of variable ratio should be given. |
| Insulation and voltage test and instructions |
terminalTX( )Terminal is same |
| The power socket can withstand the housing1.5kV,1minwithstand voltage |